Grindinės TC tipo klimatinės kameros
• Standard environmental test. chambers
• High quality 304 stainless steel exterior and interior.
• Casters for easy mobility when installing or moving environmental test chamber.
• Convenient maintenance - available to check total operating time.
• Spacious chamber volume maximum capacity of 1500L.
• Choose from 6 different model chambers based on size.
The series provide two variations in temperature range of -35 to 150? and -60 to 150,
and six capacities of 180L, 300L, 408L, 800L, 1000L, 1500L.
• Unique door lock & trim.
Vice lock type door and foam silicone packing to ensure a perfect seal.
Door open warning alarm.
• Inner door.
Optional inner glass door comes with four holes to handle samples inside the chamber.
(Except for 800L and 1500L)
• Ventilation.
Ventilation vent is located at the upper right side of the unit.
• Chart recorder. (optional)
6-point dots type recorder with digital display.
• Viewing window for observation.
A Tempered glass window provides a clear view of samples in the chamber, beneficial during testing.
Fluorescent lighting source included.
• Cable port for external probes or wires.
50mm diameter cable port comes standard on the left side of the unit.
Additional 50mm or 80mm diameter cable ports can be custom ordered.(Optional)
Control operation
• Microprocess PID control/ Auto-tuning/ Calibration. |
|||||||||||||
|
|||||||||||||
• Computer interface. (optional) |
Safety functions
• Leakage breaker for power supply.
• Over current protector.
• Overheat protectors.
• Switch-off after alarm for over heating.
• Door opening alarm.
• Vice lock door type locking mechanism.
Accessories and options
|
Viewing Window A observation window is installed ONLY on all floor type chambers. (TH/ TC series and TH-CR-270) A tempered dual windows provided a clear view of your specimen during testing. Improved lighting provided. |
![]() |
Inner door A inner glass door is provided inside the chamber door so that specimens can be observed. Applicable to some of floor type chambers. (TH-G, I-180, 300, 408L/ TC-180, 300, 408L and TH-CR-270 ONLY) All of the tabletop chambers are provided one inner glass door. |
![]() |
Temperature and Humidity recorder (digital)
|
![]() |
Access Port A through hole of 50 or 80mm dia. (2.0 or 3.2" dia.) is provided on the wall of chamber to allow electrical cables to be introduced into the chamber. A 50mm hole is provided, maximum two holes (50 or 80mm) are optional.![]() ![]() |
![]() |
Water Supply Water supply system to supply pure water for humidification.![]() ![]() ![]() |
![]() |
Shelf Wire and perforated type shelves (included shelf brackets) are made of durable stainless steel. Standard specification shelves and shelf brackets are added as required. |
![]() |
Warning Signal Light A signal light to light up when malfunction occurs. Can be located upper plate of unit. |
Cat. No. | Description | Models |
TC-G | ||
AAA8T500
|
Recorder (6 points) | ? |
AAA8T505
|
Recorder (Digital, 6 channels) | ? |
AAA8T541
|
Interface converter included a cable (5m / 16.4ft) | ? |
AAA8T542
|
Interface cable (5m / 16.4ft) | ? |
AAA80550
|
Warning signal lamp | ? |
AAA81552
|
Inner glass door (only for 180L chambers) | ? |
AAA81551
|
Inner glass door (only for 270, 300, 408L chambers) | ? |
EDA8212
|
Wire shelf for TH-G-180 / TH-I-180 / TC-G-180 * | ? |
EDA8213
|
Wire shelf for TH-G-300 / TH-I-300 / TC-G-300 * | ? |
EDA8214
|
Wire shelf for TH-G-408 / TH-I-408 / TC-G-408 * | ? |
AAA80602-1
|
Perforated shelf for TH-G-180 / TH-I-180 / TC-G-180 * | ? |
AAA80602-2
|
Perforated shelf for TH-G-300 / TH-I-300 / TC-G-300 * | ? |
AAA80602-3
|
Perforated shelf for TH-G-408 / TH-I-408 / TC-G-408 * | ? |
AAA8T610
|
Cable port (Ø 50mm / 2.0" | ? |
AAA8T611
|
Cable port (Ø 80mm / 3.2") | ? |
AAA8T621
|
Water purifier | ? |
AAA80622
|
Water cartridge tank | ? |
RTD0794
|
Humidity sensor | ? |
CFA1910
|
Temperature sensor (RTD type) | ? |
Specifications
Model | TC-G-180 | TC-G-300 | TC-G-408 | TC-G-800 | TC-G-1000 | TC-G-1500 | |
Chamber volum(L/cu ft) | 180 / 6.4 | 300 / 10.6 | 408 / 14.4 | 800 / 28.3 | 1000 / 35.3 | 1500 / 53.0 | |
Control system | Cooling fuzzy control system | ||||||
Controller | Programmable, Touch screen type controller (5.7" B/W monitor) | ||||||
Temperature | Range(?/?) | -35 to 150 / -31 to 302 | |||||
Fluctuation1)(±?/?) at 100? |
0.2 / 0.36
|
0.1 / 0.18
|
|||||
Variation1)(±?/?) at 100? |
0.9 / 1.62
|
1.0 / 1.80
|
0.8 / 1.44
|
0.4 / 0.72
|
0.5 /0.9
|
0.5 / 0.9
|
|
Time | Heating-without humidity(Min.) |
50min, -35? to 120? (-31? to 248?) |
60 From -35? to 120? (-31? to 248?) |
||||
Cooling-without humidity(Min.) |
60 From 20? to -35? (68? to -31?) |
85 From 20? to -35?
(68? to -31?) |
|||||
On single point control | |||||||
Refrigeration | System | Mechanical single stage refrigerator system (air-cooled condenser) |
Mechanical dual stage refrigerator system (air-cooled condenser) |
||||
Refrigerant |
R-404A
|
R-404A (1st step), R-23 (2nd step) |
|||||
Capacity(W) |
1500
|
4250
|
6000
|
||||
Cooler | Plate fin cooler | ||||||
Heater | Material | Ni-Cr Wire Heater |
SUS TUBE Heater
|
||||
Drying bulb(W) |
2000
|
3000
|
5000
|
||||
Temperature sensor |
Pt 100
|
||||||
Materal | Interior |
Stainless steel #304
|
|||||
Exterior |
Stainless steel #304
|
||||||
Insulation |
Polyurethane foam 100mm (3.94”)
|
||||||
Door |
Glass wool
|
||||||
Air flow | Tangent blower (90W) | 4 sets of sirocco fan (4 X 90W) | |||||
Dimensions (WxDxH) |
Interior (mm/inch) |
500x630x 600 / 19.7x24.8 x23.6 |
600x630x 800 / 23.6x24.8 x31.5 |
600x850x 800 / 23.6x33.5 x31.5 |
1000x800x 1000 / 39.4x31.5 x39.4 |
1000x1000x 1000 / 39.4x39.4 x39.4 |
1000×1000×
1500 / 39.4×39.4 ×59.1 |
Exterior (mm/inch) |
1025x1035x 1530 / 40.4x40.7 x60.2 |
1120x1070x 1730 / 44.1x42.1 x68.1 |
1120x1240x 1730 / 44.1x44.8 x68.1 |
1570x1245x 2060 / 61.8x49.0 x81.1 |
1570x1445x 2060 / 61.8x56.9 x81.1 |
1570×2045×
2050 / 61.8×80.5 ×80.7 |
|
Viewing windows (mm/inch) |
200X300 / 7.9X13.0 | 270X330 / 10.6X13.0 | |||||
Net weight (kg/lbs) |
310 ±10 / 683.4 ±22.0 |
375 ±10 / 826.7 ±22.0 |
400 ±10 / 881.8 ±22.0 |
610±10 /
1345±22.0 |
670±10 /
1477±22.0 |
750±10 /
1653.5±22.0 |
|
Cable access hole | Basically one hole provided/ Ø 50mm (2.0") Optionally max. two holes available/ Ø 50 or 80mm (2.0 or 3.2") |
||||||
Interface | Basically RS-23C [RS-485(Maximum 9 Units Control)-Optional] | ||||||
Electrical Requirements (230V, 60Hz) |
1ph / 16.2A
|
-
|
-
|
||||
Cat. No. | AAH81411 | AAH81421 | AAH81431 |